Products

XDCZ-V Laminar Flow Disinfection Canopy (Bed Isolation System)
Laminar flow disinfection canopy uses vertical laminar flow system to make the air cleanliness in the canopy maintain cleanliness 100. Simultaneously, the area inside the canopy maintains relative positive pressure to prevent the external polluted air from entering the canopy, isolating the inner from the outside. Locally clean room can be created by applying this machine. The machine can act as sterile room reducing the possibility of being infected. This is suitable for burn patient, hematological patient, patient with respiratory disease, patient receiving chemoradiotherapy and other immunocompromised patient.
Download PDF >
General Details
Specification

Disinfection principle

•UV + fiberglass HEPA

•High intensity UV can penetrate into cells of microorganisms, destroy protein structure to kill or sterilize microorganisms.

•HEPA 13 effectively eliminate particles ≥0.3 μm and small size microorganisms to achieve disinfection and cleaning effect.


Function

•After running for 10 min airborne microorganisms in the canopy is 1 cfu/m3, conforming to GB/T 16293-2010 testing method and standard. 

•After running for 10 min cleanliness is as below:

Particle Φ > 5 μm, 0/L; Particle Φ > 0.5 μm, 1/L conforming to GB 50333-2013 assessment standard. Cleanliness is class 100.


Application scope

The machine can act as sterile room reducing the possibility of being infected. This is suitable for burn patient, hematological patient, patient with respiratory disease, patient receiving chemoradiotherapy and other immunocompromised patient.


Features

•ISO 5 cleanliness can be maintained inside

•Can be run with people around.

•Precision-cut metal housing, anti-aging, anti-self-ignition.

•Vertical laminar air flow, cleanliness class 100 achieved in the canopy.

•High efficiency and low resistance HEPA 13 removes 0.5μm particles 

•Utilize high intensity UV lamp. Highly efficient killing rate, long useful life.

•Transparent, odorless, anti-static electricity curtain

•Remoter controller with magnet absorbs to housing.

•Manual, reservation modes satisfy various demands.

•Enhanced disinfection to disinfect surface in the canopy.

•Reservation mode helps to reserve 6 disinfection periods according to your requirements.

•Accumulative running time counting, automatic maintenance hint.

•LCD display panel, one click to lock to avoid misoperation

•Superior fan, low noise, stable running, adjustable air flow.


JIAJING XDCZ-V Bed Isolation System
Outline specificationsModel: XDCZ-V
Dimension: 2200(L)×1100(W)×2000(H) mm3
Technical specificationsDisinfection method: UV+HEPA
UV leakage: <1μw/cm²; authoritative report
Ozone leakage: <0.003mg/m3; authoritative report
UV intensity inside the machine ≥27000μw/cm2; authoritative report
After running machine for 1 hour, particles (≥0.5μm) is ≤300000/m³, conforming to GB 50333-2013 (Architectural technical code for hospital clean operating department). Cleanliness can achieve class10, 000.
Rate of killing bacteria in aerosol chamber >99.90%; authoritative report.
After running for 10 min airborne microorganisms in the canopy is 1 cfu/m3, conforming to GB/T 16293-2010 testing method and standard. Cleanliness in canopy can achieve Class 100. Authoritative report.
After running for 10 min cleanliness is as below:    Particle Φ > 5 μm, 0/L; Particle Φ > 0.5 μm, 1/L conforming to GB 50333-2013 assessment standard. Cleanliness is class 100. Authoritative report available.
FeaturesCan be run with people around.
Precision-cut metal housing, anti-aging, anti-self-ignition.
Vertical laminar air flow, cleanliness class 100 achieved in the canopy.
High efficiency and low resistance HEPA 13 removes 0.5μm particles
Utilize high intensity UV lamp. Highly efficient killing rate, long useful life.
Transparent, odorless, anti-static electricity curtain
Remoter controller with magnet absorbs to housing.
Manual, reservation modes satisfy various demands.
Enhanced disinfection to disinfect surface in the canopy.
Reservation mode helps to reserve 6 disinfection periods according to your requirements.
Accumulative running time counting, automatic maintenance hint.
LCD display panel, one click to lock to avoid misoperation
Superior fan, low noise, stable running, adjustable air flow.

Get a Free Quote

Submit